Product name |
RN PET5 |
Temperature |
Technique: Thermistor probe (2.25K/10k); |
Measuring and Alarm Range: 0.0℃~50℃; |
Unit: Celsius (℃); |
Resolution: 0.1℃ or 1F; |
Accuracy: ±0.1℃ (exclusive probe); |
Pulse Rate |
Measuring and Alarm Range: 25~250bpm; |
Accuracy: ±3bpm; |
Resolution: 1bpm; |
ECG |
Lead mode: 3/5 Leads, l, l, ll orl, l, l, AVR, AVL, AVF, V; |
Gain:1/4 file (x0.25)2.5mm/mV; 1/2 file(x0.50) 5.0mm/mV;1 file (x1)10mm/mV; 2 files (x2) 20mm/mV; |
Sweep speed:6.25mm/s,12.5mm/s,25mm/s; |
Baseline Recovery: < 3s after defibrillation (Monitor mode), < 1s after defibrillation (Surgery mode); |
Bandwidth: Diagnostic 0.05~100, Monitor 0.5~40Hz, Surgery 1~20Hz; |
NIBP |
Method: Pulse wave oscillation method; |
Measure mode: Manual, Auto, Continue; |
Measure interval in auto mode: 1 file (X1) 10mm/mV; 2 files (X2) 20mm/mV; 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480min; |
STAT mode cycle time: Keep 5 minutes; |
Measure and Alarm Range:
Large Animal: SYS 30~280mmHg DIA 10~220mmHg MEAN 20~240mmHg,
Small Animal: SYS 30~220mmHg DIA 10~160mmHg MEAN 20~170mmHg,
Static pressure range: 0~300mmHg,
Static pressure accuracy: 3mmHg;
|
Maximum Mean error: 5mmHg; |
Maximum Standard deviation: 8mmHg; |
Static pressure accuracy: ±3mmHg; |
Resolution: 1mmHg; |
Accuracy: Maximum Mean error ± 5mmHg, Maximum Standard deviation<8mmHg; |
Over pressure Protection: Large animal~300mmHg, Small animal~240mmHg; |
Heart Rate |
Measure range: Large animal 15~300 bpm, Small animal 15~350 bpm; |
Accuracy: 1% or 1bpm, whichever is greater; |
Resolution: 1 bpm; |
SpO2 |
Measurement Range: 0~100%; |
Resolution: 1%; |
Response Modes: Low, Medium, High; |
Accuracy: 70%~100%: 2% <69%: unspecified; |
Support Pitch tone and multi-level volume; |
EtCO2 |
Measure method: Non-dispersive infrared (NDIR); |
Measure Range: 0~19.7% (0~150mmHg) 0~20kPa; |
Resolution: 0.1 mmHg; |
CO2 Accuracy: 0~40 mmHg, ±2 mmHg, 41~70 mmHg, ±5% of reading, 71~100 mmHg, ±8% of reading, 101~150 mmHg,
±10% of reading, at 760mmHg, ambient temperature of 25℃);
|
Respiratory Rate: Range:3-150BrPM, Accuracy:±1bpm; |
Transportation and Storage |
Temperature: -20~50℃; |
Humidity: 10%~90%(non-condensing); |
Atmosphere pressure: 86KPa~106KPa; |
Respiration |
Method:lmpedance between RA-LL; |
Gain: x0.25, x0.50, x1, x2, x4; |
Respiration Rate:0~120/0-150BrPM; |
Sweep speed: 6.25 mm/s,12.5 mm/s,25mm/s; |
Resolution:1 BrPM; |
Apnea Alarm:10~40s; |
Measure range: Large animal: 0-120 bpm; |
Small animal:0-150 bpm; |
Trend & Reviewing |
NlBP measurement reviewing: 1500groups; |
Waveform review: 3 hours 600groups; |
Alarm event: 1000 groups of parameter alarms events and associated parameter; |